FDA Recall Terminated

Guidant INSIGNIA I Entra-- family of pacemakers includes the following: SSI (model numbers 0484 and 0485); DDD (models 0985, 0986); SR (models 1195, 1198); and DR (models1294, 1295, 1296). --Intermedics NEXUS Entra--family of pacemakers includes the following: SSI (model 1326); DDD (model 1426); SR (model 1398) and DR (model 1466, 1494). --The INSIGNIA I Entra pacemakers are multiprogrammable pacemakers from Guidant. -----The NEXUS I Entra pacemakers are multiprogrammable pacemakers from Intermedics. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. or Intermedics, Inc., 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Quantities Distributed by model are as follows: INSIGNIA ENTRA SSI model 484=704 devices, model 485 = 10; DDD model 985 = 13, model 986 = 779; SR model 1195 = 393, model 1198 = 2,693; DR model 1294 = 2,029, model 1295 = 439, model 1296 = 2,361; NEXUS ENTRA SSI model 1326 = 55; DDD model 1426 = 7; SR model 1395 = 45, 1398 = 117; DR model 1466 = 264, model 1494 = 46, model 1495 = 17.

Recall: Z-1293-06 · Initiated June 24, 2006

Recall

Recall Number
Z-1293-06
Event Number
35754
Firm
Boston Scientific CRM Corp
FEI Number
2124215
Product Code
LWP
Status
Terminated
Root Cause
Other
Initiated
June 24, 2006
Posted
September 28, 2006
Terminated
March 2, 2008
Address
4100 Hamline Ave N, Saint Paul, MN, 55112-5700

Description

Guidant INSIGNIA I Entra-- family of pacemakers includes the following: SSI (model numbers 0484 and 0485); DDD (models 0985, 0986); SR (models 1195, 1198); and DR (models1294, 1295, 1296). --Intermedics NEXUS Entra--family of pacemakers includes the following: SSI (model 1326); DDD (model 1426); SR (model 1398) and DR (model 1466, 1494). --The INSIGNIA I Entra pacemakers are multiprogrammable pacemakers from Guidant. -----The NEXUS I Entra pacemakers are multiprogrammable pacemakers from Intermedics. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. or Intermedics, Inc., 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Quantities Distributed by model are as follows: INSIGNIA ENTRA SSI model 484=704 devices, model 485 = 10; DDD model 985 = 13, model 986 = 779; SR model 1195 = 393, model 1198 = 2,693; DR model 1294 = 2,029, model 1295 = 439, model 1296 = 2,361; NEXUS ENTRA SSI model 1326 = 55; DDD model 1426 = 7; SR model 1395 = 45, 1398 = 117; DR model 1466 = 264, model 1494 = 46, model 1495 = 17.

Reason

Guidant has determined that low-voltage capacitors from a single component supplier may malfunction in a manner that can lead to premature battery depletion or loss of pacing output without warning in the affected devices.

Action

A Press Release was issued on 06/26/06. An Urgent Medical Device Safety Information & Corrective Action letter, dated 06/23/06 was delivered to physicians describing preliminary information before Guidant has completed its own investigation and finalized patient care recommendations. Non-implanted inventory is being retrieved from hospital shelves and Guidant Represenatatives. Physicians are to perform an in-clinic follow-up exam as soon as possible with patients implanted with affected devices.

Distribution

Worldwide- USA and Antigua Barbuda, Argentina, Australia, Austria, Belgium, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guam, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Micronesia, Netherlands, Netherlands Antilles, New Zealand, Norway, Panama, Peru, Philippines, Polynesia, Portugal, Qatar, Reunion, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan R O C, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Venezuela

Quantity

9,972