FDA Recall Terminated

VITROS ECi/ECiQ Immunodiagnostic System Product Code: 1922814 For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents

Recall: Z-1292-2019 · Initiated February 20, 2019

Recall

Recall Number
Z-1292-2019
Event Number
82311
Firm
Ortho-Clinical Diagnostics
FEI Number
1000136573
Product Code
KHO
Status
Terminated
Root Cause
Software design
Initiated
February 20, 2019
Terminated
September 15, 2021
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101

Description

VITROS ECi/ECiQ Immunodiagnostic System Product Code: 1922814 For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents

Reason

Luminometer Malfunction May Cause Inability to Process MicroWell Assays on VITROS Systems

Action

Ortho Clinical Diagnostics (Ortho) issued an "URGENT PRODUCT CORRECTION NOTIFICATION" , a customer letter (Ref. CL2019-052) dated 21 February 2019 via FedEx overnight courier and/or US Postal Service Priority Mail (for PO Boxes only) to customers who were shipped affected VITROS Systems/or whose VITROS System has a luminometer parts replacement. The customers were notified of the issue and instructed to notify Ortho upon experiencing the issue described: -Avoid performing a shutdown/restart on your systems unless directed by an Ortho representative or prompted by condition code help text. -If you shutdown/restart and encounter any of the above Luminometer condition codes and are unable to process MicroWell assays, immediately contact the Ortho Care" Technical Solutions Center for service to your system. -Complete the enclosed Confirmation of Receipt form no later than March 1, 2019 by Fax: 1.888.557.3759 or 1.585.453.4110 or e-Mail Address: [email protected]. -Please forward this notification if the product was distributed outside of your facility. Foreign affiliates were informed by email on 21 February 2019 of the issue and instructed to notify affected consignees of the issue and appropriate actions to take. Questions contact Ortho Care" Technical Solutions Center at 1-800-421-3311.

Distribution

Worldwide Distribution: US (Nationwide) and countries of: Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden,The Netherlands, and United Kingdom.

Quantity

45 units