Guidant VITALITY 2 (models T165, T175), VITALITY DS (model T125), VITALITY EL (model T135), Implantable Cardioverter Defibrillators (ICD), are designed to detect and terminate ventricular tachycardia (VT) and ventricular fibrillation (VF) and provide bradycardia therapy (atrial and ventricular pacing). Therapies include both low- and high-energy shocks using either a biphasic or monophasic waveform. Vitality 2 devices also offer a wide variety of antitachycardia pacing schemes to terminate slower, more stable ventricular tachyarrhythmias. Bradycardia pacing, including adaptive-rate features, is available to detect and treat bradyarrhythmias and to support the cardiac rhythm after defibrillation therapy. Devices denoted with DR offer dual-chamber bradycardia features (atrial and/or ventricular pacing and sensing), and the devices denoted with VR offer single-chamber bradycardia features (ventricular pacing and sensing).
Recall
- Recall Number
- Z-1292-06
- Event Number
- 35525
- Firm
- Boston Scientific CRM Corp
- FEI Number
- 2124215
- Product Code
- LWP
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 15, 2006
- Posted
- July 27, 2006
- Terminated
- March 2, 2008
- Address
- 4100 Hamline Ave N, Saint Paul, MN, 55112-5700
Description
Guidant VITALITY 2 (models T165, T175), VITALITY DS (model T125), VITALITY EL (model T135), Implantable Cardioverter Defibrillators (ICD), are designed to detect and terminate ventricular tachycardia (VT) and ventricular fibrillation (VF) and provide bradycardia therapy (atrial and ventricular pacing). Therapies include both low- and high-energy shocks using either a biphasic or monophasic waveform. Vitality 2 devices also offer a wide variety of antitachycardia pacing schemes to terminate slower, more stable ventricular tachyarrhythmias. Bradycardia pacing, including adaptive-rate features, is available to detect and treat bradyarrhythmias and to support the cardiac rhythm after defibrillation therapy. Devices denoted with DR offer dual-chamber bradycardia features (atrial and/or ventricular pacing and sensing), and the devices denoted with VR offer single-chamber bradycardia features (ventricular pacing and sensing).
Guidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.
An Urgent Medical Device Safety Information & Corrective Action letter, dated May 12, 2006, was sent to physicians. The letter gave a description of the issue, clinical implications, rate of occurrence, devices affected and recommendation. Physicians are asked to follow-up with affected patients as soon as possible because ERI to EOL time may be shortened. Physicians are asked to contact Guidant for instructions on performing a baseline 'save to disk' and provide this information to Guidant. Guidant will be able to analyze the device memory data and estimate remaining longevity and provide individualized follow-up and replacement guidelines. Subsequent follow-up visits should continue to monitor for evidence of premature battery depletion.
Nationwide including Puerto Rico. Worldwide, including Argentina, Belgium, Canada, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, New Zealand, Portugal, Spain, Syrian Arab Republic, United Arab Emirates