Corin BIOLOX DELTA MOD HEAD 36mm EX LONG +8mm12/14 TAPER. Corin Trinity Acetabular System with Extra Long Heads/Trinity Biolox Delta Ceramic Head (Generic Name: Hip Prosthesis). The Trinity Acetabular System is a modular acetabular cup system consisting of a press fit, titanium alloy shell; acetabular liners in neutral offset, +4mm offset, +4mm oblique, neutral 4mm EPW; and ceramic and CoCrMo modular heads. The ceramic and CoCrMo modular heads are compatible with Corin titanium femoral stems. The Trinity Biolox Ceramic heads are packaged in a double peel pouch configuration.
Recall
- Recall Number
- Z-1291-2021
- Event Number
- 87293
- Firm
- Corin Ltd Unit 1-57-8 Corinium Centre Cirencester United Kingdom
- FEI Number
- 3002806725
- Product Code
- MEH
- Status
- Terminated
- Root Cause
- Labeling mix-ups
- Initiated
- January 25, 2021
- Terminated
- November 23, 2022
Description
Corin BIOLOX DELTA MOD HEAD 36mm EX LONG +8mm12/14 TAPER. Corin Trinity Acetabular System with Extra Long Heads/Trinity Biolox Delta Ceramic Head (Generic Name: Hip Prosthesis). The Trinity Acetabular System is a modular acetabular cup system consisting of a press fit, titanium alloy shell; acetabular liners in neutral offset, +4mm offset, +4mm oblique, neutral 4mm EPW; and ceramic and CoCrMo modular heads. The ceramic and CoCrMo modular heads are compatible with Corin titanium femoral stems. The Trinity Biolox Ceramic heads are packaged in a double peel pouch configuration.
The size indicated on the labeling on the outer packaging, on the inner packaging, and on the patient stickers inside the packaging is incorrect: the Trinity Biolox Delta Ceramix Head 32XL is labeled as 36XL and vice-versa.
The firm contacted their consignee by email on 01/25/2021. The distributor was instructed to identify the affected product in its inventory and return it to Corin USA in order to contain the part. A recall letter will be sent and acknowledgement of receipt will be requested.
Worldwide distribution - US Nationwide distribution and the countries of Colombia, Poland, Germany, Australia.
23 units