FDA Recall Open, Classified

(1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System; (2) Lumenis Pulse 120H, Moses 2.0 (DOM), REF GA0006802, Laser System; and (3) Lumenis Pulse 120H, REF GA-0008700, Laser System.

Recall: Z-1290-2025 · Initiated January 23, 2025

Recall

Recall Number
Z-1290-2025
Event Number
96197
Firm
LUMENIS, LTD. Yokneam Ind. Park 6, Ha-Kidma Yokne'Am Ilit Israel
FEI Number
3004135191
Product Code
GEX
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
January 23, 2025
Posted
March 5, 2025

Description

(1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System; (2) Lumenis Pulse 120H, Moses 2.0 (DOM), REF GA0006802, Laser System; and (3) Lumenis Pulse 120H, REF GA-0008700, Laser System.

Reason

The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.

Action

The recall was initiated via letter dated and issued on 1/23/2025 via FEDEX in the U.S. and via mail, courier, or hand delivery OUS. The letter explained the issue and the scenarios as to when the laser system would not power on. The consignee is instructed to immediately discontinue use and segregate the device, post this information in a visible location near the affected product to ensure the information is readily accessible to all handlers and users of the device, share the notice with any health care professional within your organization and with any organization where the affected product may have been transferred, and if the device has been further distributed, notify the customer to ensure this product removal is carried out to the end-user level. A Reply Verification Tracking Form was enclosed for completion and return. Once the firm receives the completed form, the consignee will be contacted within 2 business days and provided an RGA number for product return.

Distribution

Distribution was made to PA. There was no government/military distribution. Foreign distribution was made to Mexico and Spain.

Quantity

3 devices