FDA Recall
Terminated
Depuy P.F.C. E Knee System, stabilized plus tibial insert; polyethylene tibial insert, Sz 2, 15 mm; Ref 96-2712.
Recall: Z-1289-05
·
Initiated May 26, 2005
Recall
- Recall Number
- Z-1289-05
- Event Number
- 32192
- Firm
- Depuy Orthopaedics, Inc.
- FEI Number
- 1818910
- Product Code
- NJL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 26, 2005
- Posted
- August 12, 2005
- Terminated
- December 2, 2005
- Address
- 700 Orthopaedic Dr, Warsaw, IN, 46582-3994
Description
Depuy P.F.C. E Knee System, stabilized plus tibial insert; polyethylene tibial insert, Sz 2, 15 mm; Ref 96-2712.
Reason
An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
Distribution
Nationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.
Quantity
34