FDA Recall Completed

Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (medium) C3, REF: D138502, Rx Only, CE 2797, STERILE EO, UDI: (01)10846835016277 - Product Usage: provide accessibility and maneuverability in the cardiac anatomy during cardiac electrophysiological mapping and ablation procedures.

Recall: Z-1287-2021 · Initiated February 26, 2021

Recall

Recall Number
Z-1287-2021
Event Number
87440
Firm
Biosense Webster, Inc.
FEI Number
2029046
Product Code
DYB
Status
Completed
Root Cause
Device Design
Initiated
February 26, 2021
Address
15715 Arrow Hwy, Irwindale, CA, 91706-2006

Description

Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (medium) C3, REF: D138502, Rx Only, CE 2797, STERILE EO, UDI: (01)10846835016277 - Product Usage: provide accessibility and maneuverability in the cardiac anatomy during cardiac electrophysiological mapping and ablation procedures.

Reason

There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism

Action

On 2/26/21, Biosense Webster sent an "Urgent: Medical Device Correction" notification to all affected consignees via UPS. In addition, to informing consignees about the recall, the firm asked consignees to take the following actions: 1. Precaution on inserting a dilator into VIZIGO Bi-Directional Guiding Sheath: " Use best practices for inserting or retracting any device at the hemostatic valve. " Do Not remove dilator or catheter rapidly. Damage to hemostatic valve may occur. " Aspirate slowly, only from the sideport. " Slowly remove or insert the dilator or other devices. " Once the sheath is inserted into the vasculature and the dilator is removed, aspirate until steady blood return is achieved prior to flushing or infusion. " Prior to inserting the device into the patient, pre-assemble sheath, dilator and stylet on the table. Advance the needle through the dilator and check for excessive resistance as the tip of the needle advances through the curvature of the sheath/dilator assembly. " Before inserting the sheath into the patient, flush the sheath and dilator with heparinized normal saline to remove air bubbles and any potential particulate. After the sheath is in the left atrium of the patient, maintain a constant flow of heparinized normal saline to the sheath to minimize the risk of air emboli. 2. Please review this letter carefully and share it with anyone in your facility that needs to be informed. 3. Please complete, sign, and return the Business Reply Form.

Distribution

Worldwide distribution.