FDA Recall Terminated

Whole body Computed Tomography X-Ray System (Brilliance 64 and Ingenuity CT), Brilliance 64 M/N 728231, Ingenuity CT M/N 728326, Dist. By Philips Medical Systems, Cleveland, OH

Recall: Z-1285-2012 · Initiated February 17, 2012

Recall

Recall Number
Z-1285-2012
Event Number
61363
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Software change control
Initiated
February 17, 2012
Posted
March 23, 2012
Terminated
February 22, 2013
Address
595 Miner Road, Cleveland, OH, 44143-2131

Description

Whole body Computed Tomography X-Ray System (Brilliance 64 and Ingenuity CT), Brilliance 64 M/N 728231, Ingenuity CT M/N 728326, Dist. By Philips Medical Systems, Cleveland, OH

Reason

A number of non-conformances were identified during internal and external validation testing of the Brilliance 64 and Ingenuity CT systems with software version Beta 3 (version 4.0.0.xx368).

Action

Philips Healthcare sent an Urgent Medical Device Correction letter dated February 17, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the information contained in the letter with all memebers of their staff who need to be aware of the contents in the communication. Customers were asked to retain a copy of the communication with the equipment instruction for Use. For further information or support concerning this issue, customers should contact their local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center at 1-800-722-9377, option 5. For questions regarding this recall call 440-483-7000.

Distribution

Worldwide Distribution - USA including OH and the following countries: Australia, China, Belgium, Israel & Sweden.

Quantity

7 units