FDA Recall Terminated

Talent Captivia Stent Graft System - FreeFlow configurations only. Model numbers affected have the following designation: TFXXXXXXXXCP and TBXXXXXXXXCP. Medtronic Inc. Santa Rosa, CA 95403. The Talent Thoracic Stent Graft System is designed to treat thoracic aortic aneurysms using an endovascular approach. When placed within the target lesion, the stent graft provides an alternative conduit for blood flow within a patients vasculature by excluding the lesion from blood flow and pressure.

Recall: Z-1284-2012 · Initiated February 21, 2012

Recall

Recall Number
Z-1284-2012
Event Number
61278
Firm
Medtronic Vascular, Inc.
FEI Number
3001452571
Product Code
MIH
Status
Terminated
Root Cause
Process design
Initiated
February 21, 2012
Posted
March 23, 2012
Terminated
July 12, 2012
Address
3576 Unocal Pl, Santa Rosa, CA, 95403-1774

Description

Talent Captivia Stent Graft System - FreeFlow configurations only. Model numbers affected have the following designation: TFXXXXXXXXCP and TBXXXXXXXXCP. Medtronic Inc. Santa Rosa, CA 95403. The Talent Thoracic Stent Graft System is designed to treat thoracic aortic aneurysms using an endovascular approach. When placed within the target lesion, the stent graft provides an alternative conduit for blood flow within a patients vasculature by excluding the lesion from blood flow and pressure.

Reason

Complaints received related to the inability/difficulty to release the proximal (FreeFlow) bare stent using the tip capture mechanism.

Action

Medtronic, Inc. sent an Urgent Medical Device Safety Information letter dated February 2012, to all affected customers. The letter identified the product, the problem, and the corrective action taken by Medtronic. Medtronic updated the Instructions for Use (IFU) manual. For questions regarding this recall call 707-566-1110.

Distribution

Nationwide Distribution

Quantity

1523 units