FDA Recall Open, Classified

DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B

Recall: Z-1283-2023 · Initiated February 9, 2023

Recall

Recall Number
Z-1283-2023
Event Number
91760
Firm
DeRoyal Industries Inc
FEI Number
1043214
Product Code
KDQ
Status
Open, Classified
Root Cause
Process control
Initiated
February 9, 2023
Posted
March 23, 2023
Address
200 Debusk Ln, Powell, TN, 37849-4703

Description

DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B

Reason

The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.

Action

The firm issued an URGENT DEROYAL RECALL NOTICE its consignees by email and letter beginning on 02/09/2023. The notice explained the problem and risk, and requested the following actions be taken: Identify and quarantine any affected product found in inventory. If the product has been further distributed, please notify then of this recall. Firm is seeking product return. Questions about receiving credit for return merchandise, contact Jennifer Marsh at [email protected] or by phone 865-362-4203 or email us at [email protected]. If you have questions regarding this recall please contact us at 865-362-6465 or email us at [email protected].

Distribution

US Nationwide distribution in the states of Arizona, Texas, Virginia, Pennsylvania, California, Missouri, North Carolina, Illinois, Indiana, Ohio, Arkansas.

Quantity

27,200 devices