LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004
Recall
- Recall Number
- Z-1282-2024
- Event Number
- 93958
- Firm
- Fresenius Kabi USA, LLC
- FEI Number
- 3014732157
- Product Code
- FRN
- Status
- Completed
- Root Cause
- Software design
- Initiated
- September 16, 2022
- Posted
- March 11, 2024
- Address
- 50 High St, Ste 50, North Andover, MA, 01845-2620
Description
LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004
Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022.
The issue was resolved through an update of the software to Version 5.2.1, which was installed at customer sites in September 2022. The firm did not issue a customer letter at the time. According to the firm, this issue has been resolved in all fielded product. There have been two subsequent software upgrades since Version 5.2.1 was released. Consignees are now on v 5.8.0.
US Nationwide distribution in the states of WI and NJ.
3