FDA Recall Terminated

Artis One, Interventional, Fluoroscopic, x-ray system Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis One include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for patient extremities.

Recall: Z-1282-2016 · Initiated March 11, 2016

Recall

Recall Number
Z-1282-2016
Event Number
73606
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
OWB
Status
Terminated
Root Cause
Software in the Use Environment
Initiated
March 11, 2016
Terminated
November 17, 2017
Address
40 Liberty Blvd, Malvern, PA, 19355-1418

Description

Artis One, Interventional, Fluoroscopic, x-ray system Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis One include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for patient extremities.

Reason

After importing segmentation results of the left atrium created on the Artis One system, the possibility exists for the results to be mirrored to an electro-anatomical 3D mapping system, CARTO from Biosense Webster Inc.

Action

Siemens mailed an Important Customer Safety Notice to customers to inform them of the issue and the corrective actions they plan to take to address the problem. The letter stated that Siemens service organization will contact customers to arrange a date for the installation of the software update. Customers were instructed to call Siemens service organization for an appointment at 1-800-888-7436.

Distribution

Nationwide Distribution to Michigan only

Quantity

1 angiography system