FDA Recall Terminated

Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052). The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used when locking the Femoral Nail-EX. The Aiming Arm for Titanium (TI) Cannulated Tibial Nails-EX is used when locking the Tibial Nail-EX.

Recall: Z-1281-2015 · Initiated February 17, 2015

Recall

Recall Number
Z-1281-2015
Event Number
70455
Firm
Synthes, Inc.
FEI Number
3005180112
Product Code
FZX
Status
Terminated
Root Cause
Process control
Initiated
February 17, 2015
Posted
March 16, 2015
Terminated
August 26, 2016
Address
1302 Wrights Ln E, West Chester, PA, 19380-3417

Description

Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052). The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used when locking the Femoral Nail-EX. The Aiming Arm for Titanium (TI) Cannulated Tibial Nails-EX is used when locking the Tibial Nail-EX.

Reason

For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.

Action

An Urgent Notice: Medical Device Recall, dated February 17, 2015, was sent to all direct consignees and sales consultants instructing them to take the following actions: Take note of the quantity of affected parts in your possession, and then follow ONE of the options for replacement of the affected product: Option 1: Call DePuy Synthes Repair Department at 1-800-288-6698 to schedule a repair for the affected products. Complete the Verification Section following the instructions provided and send a copy of the completed Verification Section to DePuy Synthes by: Fax: 866-723-2823 or Scan/e-mail: [email protected] Option 2: Call DePuy Synthes Repair Department at 1-800-288-6698 to order repair part. The replacement part will be sent directly to your facility. Questions should be directed to 610-719-5450 or a local DePuy Synthes sales consultant.

Distribution

Nationwide Distribution

Quantity

183