FDA Recall
Terminated
Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 18 mm diameter, 130 mm length, non-sterile, catalog no. 4309-18-13 (00-4309-018-13).
Recall: Z-1281-06
·
Initiated June 16, 2006
Recall
- Recall Number
- Z-1281-06
- Event Number
- 35783
- Firm
- Zimmer Inc.
- FEI Number
- 1000220733
- Product Code
- HWT
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 16, 2006
- Posted
- July 27, 2006
- Terminated
- December 5, 2007
- Address
- 345 E Main St, Warsaw, IN, 46580-2746
Description
Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 18 mm diameter, 130 mm length, non-sterile, catalog no. 4309-18-13 (00-4309-018-13).
Reason
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
Action
U.S. and international distributors were notified of the recall via letter dated 6/30/2006 to check customer inventories and to return recalled product to the firm. Trial implants were redesigned and reworked to eliminate the problem.
Distribution
Nationwide and worldwide including Australia, Canada, Germany, Italy and Spain.
Quantity
324 total