FDA Recall Open, Classified

Atellica IM Erythropoietin (EPO) Assay (100 Test)

Recall: Z-1280-2024 · Initiated January 24, 2024

Recall

Recall Number
Z-1280-2024
Event Number
93939
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2432235
Product Code
GGT
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
January 24, 2024
Posted
March 11, 2024
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

Atellica IM Erythropoietin (EPO) Assay (100 Test)

Reason

Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots.

Action

An Urgent Medical Device Correction notification letter dated 01/24/24 was sent to customers. Actions to be Taken by the Customer: Discontinue use of and discard the EPO kit lots listed in Table 1. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. For continued EPO patient testing in your laboratory during this time, please contact your Siemens representative to discuss alternative Siemens Healthineers solutions. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Distribution

Worldwide - US Nationwide distribution in the states of AL, CA, IA, IL, MI, MN, NC, NE, NV, NY, OH and TX; The countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, France, Germany, Greece, India, Italy, Japan, Latvia, Malaysia, Netherlands, Norway, P.R. China, Pakistan, Poland, Portugal, Republic of South Korea, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, U.A.E., United Kingdom.

Quantity

1,057 units (46 US, 1,011 OUS)