FDA Recall Open, Classified

Scan Monitor/ScanWatch, model number hwa09 Wearable, Bluetooth-connected wrist watch that records heart activity (ECG) and oxygen saturation (SpO2)

Recall: Z-1280-2023 · Initiated December 19, 2022

Recall

Recall Number
Z-1280-2023
Event Number
91523
Firm
Withings 37 Bis Rue Du General Leclerc Issy les Moulineaux CEDEX France
FEI Number
3008820594
Product Code
DPS
Status
Open, Classified
Root Cause
Software Design Change
Initiated
December 19, 2022
Posted
March 22, 2023

Description

Scan Monitor/ScanWatch, model number hwa09 Wearable, Bluetooth-connected wrist watch that records heart activity (ECG) and oxygen saturation (SpO2)

Reason

A software bug eliminated the initial ECG activation and review of the ECG under a physician. However, the device is only cleared for use under the care of a physician.

Action

Users were initially notified via push notifications in the companion app on December 19, 2022, which instructed users to update the app and activate the ECG feature under a physician. On February 10, 2023, the firm followed up with customers via Medical Device Recall letters. Customers were instructed to ensure their Health Mate app is updated to iOS 5.13.0 or higher, and to ensure their device is synchronized with the Health Mate app. Customers should follow the instructions on their Health Mate app to activate their ECG feature. Contact Information for questions: Email: [email protected] Phone number: 1-888-229-4116 (Availability Mon - Fri, 08:00 - 22:00 EST, 13:00 - 03:00 GMT, 14:00 - 04:00 CET).

Distribution

Nationwide distribution in the US. No foreign distribution.

Quantity

38,886 (US)