FDA Recall Terminated

Passport V Monitor, Multiparameter Patient Monitor(with Arrhythmia Detection and Alarms) Intended for intra hospital use under the direct supervision of a licensed healthcare practitioner.

Recall: Z-1280-2015 · Initiated January 23, 2015

Recall

Recall Number
Z-1280-2015
Event Number
70546
Firm
Mindray DS USA, Inc. dba Mindray North America
FEI Number
2221819
Product Code
MHX
Status
Terminated
Root Cause
Process control
Initiated
January 23, 2015
Posted
March 16, 2015
Terminated
April 20, 2017
Address
800 MacArthur Blvd, Mahwah, NJ, 07430-2001

Description

Passport V Monitor, Multiparameter Patient Monitor(with Arrhythmia Detection and Alarms) Intended for intra hospital use under the direct supervision of a licensed healthcare practitioner.

Reason

An issue has been identified with Passport V Monitors invasive blood pressure function (IBP) which may provide an incorrect IBP measurement

Action

Mindray DS USA Inc sent Urgent: Passport V Monitor Field Corrective Action letters dated 1/23/2015 to hospital administrators. The letters informed customers that the IBP module may provide an incorrect IBP measurement. The letter also explained the adverse effects on patients. Customers are asked to either calibrate the IBP following the instructions provided; or, contact a Mindray Service Representative, who can perform the calibration, at 1-800-288-2121. In the case of the second option, it is recommended that the monitor be removed from service until the calibration is performed. If the first option is chosen, customers are to notify a Service Representative upon completion of the action. Customers can direct their questions to Diane Arpino at 201-995-8407.

Distribution

Nationwide Distribution

Quantity

519 units