FDA Recall Terminated

Oncomine Dx Target Test User Guides and Assay Definition File, Model: A32461; UDI: (01)10190302006071(11)170911(10)1.8(240)A32461 The Oncomine Dx Target Test is a qualitative in vitro diagnostic test that uses targeted high- throughput, parallel-sequencing technology to detect single-nucleotide variants (SNVs) and deletions in 23 genes from DNA and fusions in ROS1 from RNA isolated from formalin-fixed, paraffin-embedded (FFPE) tumor tissue samples from patients with non-small cell lung cancer (NSCLC) using the Ion PGM Dx System. The test is indicated to aid in selecting NSCLC patients for treatment with the targeted therapies listed in Table 1 in accordance with the approved therapeutic product labeling.

Recall: Z-1279-2018 · Initiated December 4, 2017

Recall

Recall Number
Z-1279-2018
Event Number
79437
Firm
Life Technologies Corporation
FEI Number
3003335080
Product Code
PQP
Status
Terminated
Root Cause
Device Design
Initiated
December 4, 2017
Terminated
November 10, 2020
Address
7335 Executive Way, Frederick, MD, 21704-8354

Description

Oncomine Dx Target Test User Guides and Assay Definition File, Model: A32461; UDI: (01)10190302006071(11)170911(10)1.8(240)A32461 The Oncomine Dx Target Test is a qualitative in vitro diagnostic test that uses targeted high- throughput, parallel-sequencing technology to detect single-nucleotide variants (SNVs) and deletions in 23 genes from DNA and fusions in ROS1 from RNA isolated from formalin-fixed, paraffin-embedded (FFPE) tumor tissue samples from patients with non-small cell lung cancer (NSCLC) using the Ion PGM Dx System. The test is indicated to aid in selecting NSCLC patients for treatment with the targeted therapies listed in Table 1 in accordance with the approved therapeutic product labeling.

Reason

The Oncomine Dx Target Test may report erroneous results when used off-label for samples other than non-small cell lung cancer (NSCLC) tumor specimens.

Action

The following actions are described in the consignee later: Review your testing results from the Oncomine Dx Target Test to ensure that none of the reports containing BRAF V600E positive samples are impacted. Utilize Figure 1 and 2 to aid in identifying this discrepancy. If you have any questions, please contact 408.802.2341.

Distribution

US Distribution to the states of: AZ, CA, NC, TX, and VA.

Quantity

9