FDA Recall Terminated

Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 This Philips MultiDiagnost Eleva is intended to be used as a multifunctional/universal system. General RIF, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications. This includes the following general areas: Digestive system, Skeletal system, Reproductive system, Respiratory system, Circulatory system, Various:Arthrograms, Myelograms, Facet joint injections, Discography, Sialography

Recall: Z-1279-2016 · Initiated September 24, 2014

Recall

Recall Number
Z-1279-2016
Event Number
69339
Firm
Philips Medical Systems, Inc.
FEI Number
1218950
Product Code
KPR
Status
Terminated
Root Cause
Device Design
Initiated
September 24, 2014
Terminated
March 26, 2019
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 This Philips MultiDiagnost Eleva is intended to be used as a multifunctional/universal system. General RIF, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications. This includes the following general areas: Digestive system, Skeletal system, Reproductive system, Respiratory system, Circulatory system, Various:Arthrograms, Myelograms, Facet joint injections, Discography, Sialography

Reason

Loss of key image functionality due to a bent pedal of the Footswitch.

Action

Field Safety Noticed was mailed on September 24, 2014 via Certified Return Receipt. This FCC consists of a Mechanical Hardware update of the Footswitch in order to prevent bent pedals and therefore prevents an intermittent or continuous inability of making live fluoro images or exposures. Customers with questions were instructed to contact their local Philips representative.

Distribution

Worldwide Distribution - US (nationwide) and Internationally to: CZECH REPUBLIC ALBANIA ALGERIA ARGENTINA AUSTRALIA AUSTRIA AZARBAIJAN BAHAMAS BANGLADESH BELGIUM BRAZIL BULGARIA CAYMAN ISLANDS CHILE CHINA COLUMBIA COSTA RICA CROATIA CUBA CZECHOSLOVAKIA DENMARK DOMINICAN REPUB ECUADOR EGYPT ESTONIA FINLAND FRANCE GABON GERMANY GREECE HONG KONG HUNGARY ICELAND INDIA INDONESIA IRAN IRAQ IRELAND ISRAEL ITALY JAPAN JORDAN KAZAKHSTAN KENYA KOREA SOUTH KUWAIT LATVIA LEBANON LIBYA LITHUANIA MACEDONIA MALAYSIA MAURITIUS MEXICO MOLDAVIA MONGOLIA MOROCCO NEPAL NETHERLANDS NETHERLANDS ANT NEW ZEALAND NORWAY OMAN PAKISTAN PALESTINE PANAMA Papua New Guinea PARAGUAY PERU PHILIPPINES POLAND PORTUGAL PUORTO RICO QATAR REUNION ROMANIA RUSSIAN FEDERAT SAUDI ARABIA SERBIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SUDAN SWEDEN SWITZERLAND SYRIA TAIWAN TANZANIA THAILAND TUNISIA TURKEY UKRAINE UNITED ARAB EMI United Arab Emirates UNITED KINGDOM URUGUAY USA UZBEKISTAN VENEZUELA VIETNAM YEMEN GUADELOUPE SYRIAN ARAB REPUBLIC

Quantity

168