FDA Recall Terminated

Codman ICP Skull Bolt Kit, Product Code: 82-6638. Single Use, disposable drill kit for gaining access to the cranium to allow for insertion and retention of intracranial cranial pressure sensor, included in the kit.

Recall: Z-1279-2012 · Initiated February 9, 2012

Recall

Recall Number
Z-1279-2012
Event Number
61205
Firm
Codman & Shurtleff, Inc.
FEI Number
3002947440
Product Code
GWM
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
February 9, 2012
Posted
March 22, 2012
Terminated
September 11, 2015
Address
325 Paramount Drive, Raynham, MA, 02767

Description

Codman ICP Skull Bolt Kit, Product Code: 82-6638. Single Use, disposable drill kit for gaining access to the cranium to allow for insertion and retention of intracranial cranial pressure sensor, included in the kit.

Reason

Codman ICP Skull Bolt wrench is out of specification and cannot be used to turn the screw on the drill

Action

Codman issued "Urgent - Voluntary Product Recall Notification" letters on 2/9/12 via Fedx in the US and requested accounts examine inventory and return product. An Acknowledgement Form was requested to be returned. Outside the US affiliates were notified to recall to their accounts. Questions were directed to Customer Service (800) 225-0460.

Distribution

Worldwide Distribution - USA, and the countries of Canada, Austria, Brazil, Belgium, Columbia, China, Denmark, Estonia, Spain, Finland, France, Great Britain, Germany, India, Italy, Japan, Lithuania, Netherlands, New Zealand, Oman, Mexico, Portugal, Russia, Singapore, SIovenia, and Switzerland.

Quantity

4056 kits