Estradiol III; Elecsys and cobas e analyzers; REF 06656021 190; M 6.5 mL, R1 9 mL, R2 9 mL, For USA: CONTENT M Streptavidin-coated microparticles 0.72 mg/ml; R1 Two biotinylated monoclonal anti-estradiol antibodies (rabbit) 2.5 ng/m and 4.5 ng/mL Mesterolone 130 ng/mL; R2 Estradiol derivative, labeled with ruthenium complex 4.5 ng/mL; Rx only Product Usage: Immunoassay for the in vitro quantitative determination of estradiol in human serum and plasma. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.
Recall
- Recall Number
- Z-1276-2016
- Event Number
- 73411
- Firm
- Roche Diagnostics Operations, Inc.
- FEI Number
- 1823260
- Product Code
- CHP
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- March 7, 2016
- Terminated
- January 10, 2018
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025
Description
Estradiol III; Elecsys and cobas e analyzers; REF 06656021 190; M 6.5 mL, R1 9 mL, R2 9 mL, For USA: CONTENT M Streptavidin-coated microparticles 0.72 mg/ml; R1 Two biotinylated monoclonal anti-estradiol antibodies (rabbit) 2.5 ng/m and 4.5 ng/mL Mesterolone 130 ng/mL; R2 Estradiol derivative, labeled with ruthenium complex 4.5 ng/mL; Rx only Product Usage: Immunoassay for the in vitro quantitative determination of estradiol in human serum and plasma. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.
Due to the risk of a recently identified cross reactivity, the Elecsys Estradiol assay should not be used when monitoring patient being treated with fulvestrant. Customers should be informed that fulvestrant will increase the apparent concentration of estradiol in women being treated with this drug. If treatment with fulvestrant has been altered or discontinued as a result of falsely elevated e
Roche sent an Urgent Medical Device Correction letter dated March 7, 2016 via UPS Ground (receipt required) to the consignee list. The letter identified the affected product, problem and actions to be taken. Consignees were instructed to complete the attached fax form and fax it to 1-866-891-1984. For questions contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336.
US Nationwide Distribution including Puerto Rico.
23,630