FDA Recall Terminated

Osteonics Shoulder Glenoid Spanner Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430. The Solar Upper Extremity System Hemi and Total Shoulder Arthroplasty system is designed to address the most common arthritic disorders affecting the shoulder. The spherical reamer is used to prepare the glenoid fossa and provide a uniform surface for instrumentation and implant placement. The glenoid fossa is then sized using surface templates to confirm the appropriate coverage, size and curvature has been determined. To achieve this in surgery, the spherical reamer (5900-2050/-2051/-2052/-2053) must be attached to the Angled Glenoid Driver (5900-2040) by threading the reamer's male threads into the driver's female threads. The spherical reamers come in sizes 5, 7, 9 & 11, each with a captive drill head screw (5900-2054/-2055). The spanner features two kidney shaped male lobes that mate with two similar kidney shaped female slots in the screw's head and is used to tighten and loosen the spherical reamers with the angled driver. These instruments are part of the Solar Total Shoulder Instrument set and can be found in Surgical Protocol LSPUE-4 dated 10/04. Each set contains one spanner that universally fits all size reamers.

Recall: Z-1276-2012 · Initiated November 17, 2011

Recall

Recall Number
Z-1276-2012
Event Number
61312
Firm
Stryker Howmedica Osteonics Corp.
FEI Number
3003070421
Product Code
LXH
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
November 17, 2011
Posted
March 22, 2012
Terminated
December 19, 2012
Address
325 Corporate Drive, Mahwah, NJ, 07430

Description

Osteonics Shoulder Glenoid Spanner Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430. The Solar Upper Extremity System Hemi and Total Shoulder Arthroplasty system is designed to address the most common arthritic disorders affecting the shoulder. The spherical reamer is used to prepare the glenoid fossa and provide a uniform surface for instrumentation and implant placement. The glenoid fossa is then sized using surface templates to confirm the appropriate coverage, size and curvature has been determined. To achieve this in surgery, the spherical reamer (5900-2050/-2051/-2052/-2053) must be attached to the Angled Glenoid Driver (5900-2040) by threading the reamer's male threads into the driver's female threads. The spherical reamers come in sizes 5, 7, 9 & 11, each with a captive drill head screw (5900-2054/-2055). The spanner features two kidney shaped male lobes that mate with two similar kidney shaped female slots in the screw's head and is used to tighten and loosen the spherical reamers with the angled driver. These instruments are part of the Solar Total Shoulder Instrument set and can be found in Surgical Protocol LSPUE-4 dated 10/04. Each set contains one spanner that universally fits all size reamers.

Reason

Two product complaints reported that the Glenoid Reamer Spanner (Lot F3W7017) would not properly connect to the Glenoid Spherical Reamer.

Action

Stryker sent an "URGENT PRODUCT RECALL" letter dated December 21, 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Product Recall Acknowledgment Form was attached for customers to complete and return via fax to 201-831-6069. Contact the firm at 201-972-2100 for questions regarding this notice.

Distribution

Worldwide Distribution-USA (nationwide) and the country of Canada.

Quantity

five units