FDA Recall Open, Classified

smiths medical Pneupac paraPAC plus 300 ventilator kit without PEEP and CPAP, REF P300NUS

Recall: Z-1275-2024 · Initiated February 7, 2024

Recall

Recall Number
Z-1275-2024
Event Number
94089
Firm
Smiths Medical ASD Inc.
FEI Number
3012307300
Product Code
BTL
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
February 7, 2024
Posted
March 19, 2024
Address
6000 Nathan Ln N, Minneapolis, MN, 55442-1690

Description

smiths medical Pneupac paraPAC plus 300 ventilator kit without PEEP and CPAP, REF P300NUS

Reason

When a paraPAC plus ventilator is switched to the operating mode of VENTILATE, the ventilator may intermittently provide continuous positive gas flow instead of the intended cycling like a human breath. This non-cycling and continuous positive gas flow when in the cycling mode, is a malfunction, not allowing the ventilator to properly function as designed.

Action

Smiths medical issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 02/07/2024 via letter using traceable means. The notice explained the issue with the device, potential risk to the patient, and requested the following actions be taken: "There is no need to return or discontinue using your paraPAC plus, at this time. When using the device, all instructions, including warnings and cautions in the User Manual Doc. numbers (10018833-003 and/or 10026347-002) must be followed with heightened awareness. This is inclusive, but not limited to the following: " Constant monitoring of the patient. " Blood oxygenation and expired carbon dioxide levels should be monitored independently using pulse oximetry and capnography. " All pre-use checks must be performed before each use. " Alternative means of ventilation such as bag mask ventilation, must be available in the event of ventilator failure or malfunction. If the paraPac plus ventilator experiences continuous flow, remove the ventilator from clinical use, set the device aside for repair and use another device or alternative means of ventilation." " Customer Required Actions: 1. Please identify all paraPAC plusTM units in your possession. 2. Share this recall notification with all potential users of the devices and if the devices are used at another location. 3. Complete and return the attached Customer Response Form to [email protected]. 4. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]." For questions about this Field Correction Notice, contact your sales representative or call 763-383-3072.

Distribution

Worldwide distribution: US (nationwide) and OUS countries of: Algeria, Armenia, Australia, Austria, Bangladesh, Bahrain, Bolivia, Plurinational state of, Brazil, Bulgaria, Canada, Colombia, Czech Republic, Cyprus, Ecuador, Egypt, El Salvador, Estonia, France, Germany, Greece, Hong Kong, Iceland, India, Indonesia, Israel, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Singapore, Slovakia, Slovenia, Saudi Arabia, South Africa, Spain, Sweden, Tawain, Province of China, Thailand, Trinidad and Tobago, Turkey, United Kingdom and Viet Nam.

Quantity

2902 units