FDA Recall Terminated

Ray Station 4.9 Ray Station 5, Ray Station 6 and Ray Station 7 Software build numbers: 4.9.0.42, 5.0.0.37, 5.0.1.11, 5.0.2.35, 6.0.0.24, 6.1.0.26, 6.1.1.2, 6.2.0.7 or 7.0.0.19 UDI: 0735000201006820171130

Recall: Z-1275-2018 · Initiated January 26, 2018

Recall

Recall Number
Z-1275-2018
Event Number
79582
Firm
RAYSEARCH LABORATORIES AB Sveavaegen 9 Stockholm Sweden
FEI Number
3007774465
Product Code
MUJ
Status
Terminated
Root Cause
Device Design
Initiated
January 26, 2018
Terminated
January 13, 2020

Description

Ray Station 4.9 Ray Station 5, Ray Station 6 and Ray Station 7 Software build numbers: 4.9.0.42, 5.0.0.37, 5.0.1.11, 5.0.2.35, 6.0.0.24, 6.1.0.26, 6.1.1.2, 6.2.0.7 or 7.0.0.19 UDI: 0735000201006820171130

Reason

Software issue with Center Beam in Field functionality. Issue can result in incorrect treatment volume delivered to patient.

Action

On January 26, 2018 RaySearch Laboratories AB sent Field Safety Notice, Medical Device Correction #25955 to each customer via e-mail. Customers were advised of the following: be aware that Center Beam in Field affects all beams that share the same isocenter and that apertures may be modified in a way not intended; do not use Center Beam in Field after having created an opposed beam. If an opposed beam has been created, delete the opposed beam and use Center Beam in Field on the remaining beam. Then recreate the opposed beam; please educate planning staff and all users about this workaround; inspect your product and identify all installed units with the above software version number(s), then confirm you have read and understood this notice by replying to the notification via email, fax or by calling local support. Customers with questions were instructed to contact David Hedfors via e-mail at [email protected]

Distribution

Nationwide

Quantity

2484