TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone.

Recall: Z-1273-2020 · Initiated December 30, 2019

Recall

Recall Number
Z-1273-2020
Event Number
84748
Firm
Smith & Nephew, Inc.
FEI Number
1219602
Product Code
MBI
Status
Terminated
Root Cause
Package design/selection
Initiated
December 30, 2019
Terminated
May 11, 2021
Address
130 & 120 Forbes Blvd, Mansfield, MA, 02048

Description

TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone.

Reason

Due to a potential for sterile barrier breach. A complaint was received that indicated the protective tube of the device came off inside the pouch enabling the pointed end of the device to puncture the package causing a breach of the sterile barrier, which could affect the sterility of the device.

Action

Smith Nephew "Urgent Medical Device Recall Notice R-2019-22" letter dated December 30, 2019. Customers were advised to: 1.Please inspect your inventory and locate any devices from the listed product and batch numbers on the first page of this Field Action Notification, and quarantine them immediately. a. If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to [email protected] or fax to 901-566-7975. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact [email protected].

Distribution

Worldwide distribution US Nationwide distribution in the states of AZ, ID, MO, NC, ND, NE, OK, SC, TN, and countries of AE, AU, CA, CH, ES, FI, JP, PR, SG.

Quantity

82 distributed throughout US:AZ, ID, MO, NC, ND, NE, OK, SC, TN. 243 distributed OUS: AE, AU, CA, CH, ES, FI, JP, PR, SG. Total: 325 units