Philips HeartStart FRx AED, Model 861304, FRx codeRev: E.02.017

Recall: Z-1271-2020 · Initiated September 10, 2019

Recall

Recall Number
Z-1271-2020
Event Number
84850
Firm
Philips North America LLC
FEI Number
1000524572
Product Code
MKJ
Status
Terminated
Root Cause
Process control
Initiated
September 10, 2019
Terminated
April 29, 2024
Address
22100 Bothell Everett Hwy, Bothell, WA, 98021-8431

Description

Philips HeartStart FRx AED, Model 861304, FRx codeRev: E.02.017

Reason

Affected devices contain a Printed Circuit Assembly (PCA) that was not manufactured/repaired in accordance to quality standards.

Action

On 09/11/2019, the firm sent an "Customer Information Medical Device Recall" notification to customers via Certified Mail informing them that certain models of their Automated External Defibrillator (AED) contain a Printed Circuit Assembly (PCA) manufactured by a supplier that did not follow specific procedures when repairing PCAs that initially failed required testing. After being repaired, these PCAs passed all testing at the supplier's facility prior to being assembled into AEDs. The Recalling Firm's quality standards require that the PCAs both be repaired according to specified procedures and pass testing. The Recalling Firm states that the serial number of the AED is printed on the label on the back of the device. The Recalling Firm will replace the affected devices at no charge and while awaiting for the replacement AED, the customer may continue to use the device. The Recalling Firm or distributors will contact customers to arrange for replacement and removal of affected devices. For any additional questions/information, consumers can contact the Recalling Firm at (800)263-3342 option 5 to speak with a AED Recall Team or email [email protected]

Distribution

HeartStart FRx: US - CA, CT, FL, IL, MO, NY,TN, TX and WI OUS - Australia, Brazil, Canada, Germany, Italy, Japan, South Korea, Norway, Switzerland and Taiwan. HeartStart HS1: US - CT, FL, IL, MI, MS, NJ, TN, TX, WA and WI OUS - Argentina, Australia, Canada, France, Germany, Hong Kong, Italy, Japan, Netherlands, Norway, Spain, Sweden, and Taiwan

Quantity

44 units