FDA Recall Terminated

Ellex Medical Fixed eye Safety filters Model Integre S LP5532. The filter is located in the binocular mount of the laser delivery head. The filter prevents transmission of treatment laser light to physician eyes while the aiming beam to remain visible. .

Recall: Z-1270-2016 · Initiated February 1, 2016

Recall

Recall Number
Z-1270-2016
Event Number
73468
Firm
Laserex Systems Inc.
FEI Number
3002806902
Product Code
HQF
Status
Terminated
Root Cause
Device Design
Initiated
February 1, 2016
Terminated
October 27, 2016
Address
7138 SHADY OAK RD, EDEN PRAIRIE, MN, 55344-3517

Description

Ellex Medical Fixed eye Safety filters Model Integre S LP5532. The filter is located in the binocular mount of the laser delivery head. The filter prevents transmission of treatment laser light to physician eyes while the aiming beam to remain visible. .

Reason

The Firm discovered that there is a design error which is traceable back to 2012 when a design change on the safety filter optic adaptor was implemented.

Action

Ellex Medical Pty Ltd. Planned Action to bring into Compliance: 1.Field Corrective Action replacing all affected safety filters of all units manufactured between July 2012 and February 2016, 2.All repairs will be made at no cost to the purchaser; and 3.The corrections will be completed by March 30, 2017. CDRH approves the CAP subject to the following conditions: Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. For further questions please call (800) 824-7444.

Distribution

Worldwide Distribution - US Distribution

Quantity

72 units plus 2 spare units