FDA Recall Open, Classified

Atellica CH Total Protein II- In vitro diagnostic use in the quantitative determination of total protein in human serum and plasma (lithium heparin) using the Atellica CH Analyzer. SMN: 11097604

Recall: Z-1269-2023 · Initiated January 27, 2023

Recall

Recall Number
Z-1269-2023
Event Number
91665
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2432235
Product Code
CEK
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
January 27, 2023
Posted
August 29, 2023
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

Atellica CH Total Protein II- In vitro diagnostic use in the quantitative determination of total protein in human serum and plasma (lithium heparin) using the Atellica CH Analyzer. SMN: 11097604

Reason

Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results

Action

Siemens Issued Urgent Field Notice to US customers ACHC23-01.B.US via FedEx beginning on 01/30/2023. In accordance with outside US country regulations, Siemens in-country personnel for the EU and non-EU regions follow local requirements for contacting their respective National Competent Authorities customers for distribution. and regional reporting to their local competent authorities beginning on 01/27/2023. The UMDC/UFSN provides instructions to all Atellica CH 930 customers to either separate the assays causing carryover by analyzer, if they have multiple analyzers, or by batch testing of Mg. In addition, the UMDC and UFSN indicates that SW v1.27 and higher will correct this issue. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director. Perform the instructions provided in the Additional Information section. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Distribution

Nationwide Foreign: Country Albania Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile China Colombia Croatia Cura¿ao Czech Republic Denmark Ecuador Egypt Finland France Germany Greece Guadeloupe Hong Kong Hungary India Ireland Israel Italy Japan Kuwait Libya Malaysia Mexico Netherlands New Zealand Norway Oman Pakistan Paraguay Peru Philippines Poland Portugal Qatar Republic Korea Romania Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan, Province of China Thailand Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vatikancity Vietnam

Quantity

13071 units