FDA Recall
Terminated
EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for each calibration level, and an insert sheet.
Recall: Z-1267-2018
·
Initiated April 11, 2017
Recall
- Recall Number
- Z-1267-2018
- Event Number
- 77160
- Firm
- Medica Corporation
- FEI Number
- 1220972
- Product Code
- JIT
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- April 11, 2017
- Terminated
- August 20, 2019
- Address
- 5 Oak Park Dr, Bedford, MA, 01730-1430
Description
EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for each calibration level, and an insert sheet.
Reason
Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.
Action
Consignees were notified of the recall via email bulletin dated 4/11/17. The bulletin informs customers of the issues and explains the actions to be taken.
Distribution
USA: MI FL CA TX India, Russia, Puerto Rico, N. Mariana Island/Palau.