syngo Imaging model number 10014063. Affected systems are V30A or earlier with an NFS connected archive and V30A systems where image data was stored to LTS using software version older than V30A. Picture Archiving and communications system

Recall: Z-1266-2009 · Initiated February 9, 2009

Recall

Recall Number
Z-1266-2009
Event Number
51563
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
February 9, 2009
Posted
April 28, 2009
Terminated
September 23, 2009
Address
51 Valley Stream Pkwy, Malvern, PA, 19355

Description

syngo Imaging model number 10014063. Affected systems are V30A or earlier with an NFS connected archive and V30A systems where image data was stored to LTS using software version older than V30A. Picture Archiving and communications system

Reason

An issue concerning archived image data that is sporadically saved with corrupted data may occur. This issue may occur when syngo Imaging is configured to archive image data on a network file system (NFS) connected via a connected archive (Mount Point Solution).

Action

A Customer Safety Advisory Notice, dated 02/09/2009 was sent to affected customers via Update Instruction IM013/09S. The letter informs customers of when the malfunction occurs and the potential issue, provides instructions to avoid its occurrence, and how the issue will be resolved. Customers are to ensure that the Safety Advisory is placed in the system's instructions for use and contact their SIEMENS Uptime Service Senter if they have any questions.

Distribution

Nationwide Distribution -- including states of AL, CA, CT, FL, LA, MI, MO, NC, NJ, OH, PA, TN, and WI.

Quantity

20 units