FDA Recall Terminated

Halogen Lamp component in RetCam 3 System - Clarity Medical Systems, Inc; Pleasanton, CA. General ophthalmic imaging including retinal, corneal and external imaging. Photo-documentation of pediatric ocular diseases including retinopathy of prematurity. Screening for Type-2 re-threshold retinopathy of prematurity or treatment requiring ROP.

Recall: Z-1265-2014 · Initiated March 7, 2014

Recall

Recall Number
Z-1265-2014
Event Number
67702
Firm
Clarity Medical Systems Inc
FEI Number
3004014430
Product Code
HKI
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
March 7, 2014
Posted
March 20, 2014
Terminated
August 11, 2014
Address
5775 W Las Positas Blvd, Suite 200, Pleasanton, CA, 94588-4084

Description

Halogen Lamp component in RetCam 3 System - Clarity Medical Systems, Inc; Pleasanton, CA. General ophthalmic imaging including retinal, corneal and external imaging. Photo-documentation of pediatric ocular diseases including retinopathy of prematurity. Screening for Type-2 re-threshold retinopathy of prematurity or treatment requiring ROP.

Reason

One lot of Halogen lamps have the potential to fail sooner than the expected life of the lamp which could result in damage limited to the internal electrical components inside the RetCam 3 computer module.

Action

The firm, Clarity, sent a "Field Bulletin" dated March 12, 2014 to customers - risk managers and distributors by e-mail on March 12, 2014. They plan to follow this with a second letter on or about March 28, 2014 which will include a second bulletin, replacement bulbs and installation instructions. Clarity stated in the letter to its customers "In the meantime, you may continue to use your RetCAM 3." If you have any question of concerns, please contact Clarity at +1 925 621 322 or email to: [email protected]

Distribution

Worldwide Distribution: US (nationwide) to states of: TN, MA, CA, TX, MD,FL, MS, IL, and PA; and to countries of: France, Thailand, Egypt, Mexico, Kuwait, India, Austria,Japan, Sweden, Peru, United Kingdom, Philippines,The Netherlands, China,, UAE,Israel, Germany, Italy.

Quantity

234 lamps