NM/CT 870 CZT, Model/Catalogue/Code # H3906CW Product Usage: The GE Discovery NM/CT 670 CZT system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography system which is intended for enabling attenuation correction and anatomical localization of SPECT images and for standalone head, whole body, cardiac and vascular X-ray Computed Tomography applications.
Recall
- Recall Number
- Z-1264-2019
- Event Number
- 82582
- Firm
- GE Healthcare, LLC
- FEI Number
- 2126677
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- March 20, 2019
- Terminated
- October 5, 2020
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
NM/CT 870 CZT, Model/Catalogue/Code # H3906CW Product Usage: The GE Discovery NM/CT 670 CZT system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography system which is intended for enabling attenuation correction and anatomical localization of SPECT images and for standalone head, whole body, cardiac and vascular X-ray Computed Tomography applications.
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
GE Healthcare sent an Urgent Medical Device Correction letter dated March 20, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. GE Healthcare will inspect and, if required, correct all affected products at no cost to you to ensure that all design redundancies are in place. A GE Healthcare representative will contact you to arrange for the correction. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
16 units