FDA Recall Terminated

Carina Home Ventilator. The device is used as a continuous respirator/ventilator for home use.

Recall: Z-1263-2009 · Initiated January 21, 2009

Recall

Recall Number
Z-1263-2009
Event Number
51447
Firm
Draeger Medical, Inc.
FEI Number
2517967
Product Code
CBK
Status
Terminated
Root Cause
Software design
Initiated
January 21, 2009
Posted
May 14, 2009
Terminated
January 5, 2010
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Carina Home Ventilator. The device is used as a continuous respirator/ventilator for home use.

Reason

The device may not generate a pilot line disconnect alarm if the pilot line becomes disconnected, which may potentially result in a substantial reduction in tidal volume delivery.

Action

An "Urgent - Medical Device Recall" Letter dated January 2009 was issued to consignees via certified mail. The letter described the issue, specific conditions, background information, and precautions for customers. An additional letter dated January 22, 2009 was included with the "Urgent - Medical Device Recall" letter which instructed consignees to inform an enduser of the contents of the notification letter because some of the consignees are home care device/service providers. When the new hardware/software solution is available, affected devices will be updated free of charge. Direct questions about the notification letter to Drager Medical, Inc. by calling 1-800-543-5047 (press "1" at the prompt and then press "2349#"). Direct questions regarding updating your device when the new software is available, contact Drager Service Technical Support at 1-800-543-5047 (press "4" at the prompt).

Distribution

Nationwide Distribution -- states of CA, FL, GA, HI, IL, LA, MA, MN, OH, and TX.

Quantity

40 units