FDA Recall Open, Classified

MiniMed insulin pump, REF: 700G (MMT-1801, MMT-1805, MMT-1850, MMT-1851)

Recall: Z-1262-2025 · Initiated January 31, 2025

Recall

Recall Number
Z-1262-2025
Event Number
96200
Firm
Medtronic MiniMed, Inc.
FEI Number
3003166194
Product Code
OZP
Status
Open, Classified
Root Cause
Device Design
Initiated
January 31, 2025
Posted
March 3, 2025
Address
18000 Devonshire St, Northridge, CA, 91325-1219

Description

MiniMed insulin pump, REF: 700G (MMT-1801, MMT-1805, MMT-1850, MMT-1851)

Reason

All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

Action

On January 31, 2025, Medtronic MiniMed issued a "Urgent Medical Device Correction" notification to all affected consignees. The notification was distributed via: e-mail, certified mail, SMS text message or phone call. 1. Monitor Your Glucose Levels: 2. Discuss how to prepare for situations like this with your healthcare professional. 3. Respond to Alerts and Symptoms. 4. Please acknowledge that you have read and understood this notification and have followed the actions listed in this letter by either completing and returning the confirmation form, scanning the QR code below with your phone or tablet, or by visiting [https://info.medtronicdiabetes.com/pump-delivery-volume] to acknowledge this communication.

Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY and the countries of Algeria, Andorra, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Plurinational State Of Bosnia, Herzegovina , Botswana, Brazil, Brunei, Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Cayman, Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Gibraltar, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Islamic, Republic Of Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Malawi, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Republic Of Montenegro, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, North Macedonia, Northern Ireland, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion Romania, Russia, Saint Pierre And Miquelon, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Tanzania, United Republic Of Thailand, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian Republic Of Vietnam, Yemen, Zimbabwe.

Quantity

4,959 unit