FDA Recall Terminated

Model #842-326 (pump tube for waste) & Model #842-327 (pump tube for solution), Waste Tubing used on ABL 7xx and ABL8xx, component of the ABL700/800 Blood Gas Analyzers, also included in service kits part no. 905-671

Recall: Z-1260-06 · Initiated May 22, 2006

Recall

Recall Number
Z-1260-06
Event Number
35695
Firm
Radiometer America Inc
FEI Number
1523456
Product Code
CHL
Status
Terminated
Root Cause
Other
Initiated
May 22, 2006
Posted
July 26, 2006
Terminated
April 17, 2012
Address
810 Sharon Dr, Westlake, OH, 44145-1598

Description

Model #842-326 (pump tube for waste) & Model #842-327 (pump tube for solution), Waste Tubing used on ABL 7xx and ABL8xx, component of the ABL700/800 Blood Gas Analyzers, also included in service kits part no. 905-671

Reason

An increased failure rate was observed with the yellow rubber tubing. The tubing was found to be leaking blood and solutions after distribution.

Action

The recalling firm issued a letter to their customers dated 5/22/2006. The letter informs the customer(s) of the tubing defect and provides the customer with one (1) tubing to deal with any immediate needs which they may have. Should the customer require additional waste pump tubing replacements, the letter is accompanied by a FAX FORM which can be used to request any amount of additional replacement tubing required. The letter also shows tthe address to which the replacement tubing is to be delivered, and provides a Purchase Order request to be used, if required.

Distribution

Nationwide to user/customers in the U.S. and in Canada

Quantity

2500 Units in U.S. & 76 Units in Canada