FDA Recall Terminated

Cytomics FC 500 Flow Cytometry System Part Numbers 626553 (TN; FC500 with UPS), 6605627 (TN; 5 CLR, FC500 (100V)), 6605628 (TN; 5 CLR, FC500 (120V)) , 6605629 (TN; 5 CLR, FC500 (220V)), 6605630 (TN; 5 CLR, FC500 (240V)) with CXP Software Versions 2.0 and 2.1

Recall: Z-1259-06 · Initiated May 22, 2006

Recall

Recall Number
Z-1259-06
Event Number
35643
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
LXG
Status
Terminated
Root Cause
Other
Initiated
May 22, 2006
Posted
July 26, 2006
Terminated
April 26, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

Cytomics FC 500 Flow Cytometry System Part Numbers 626553 (TN; FC500 with UPS), 6605627 (TN; 5 CLR, FC500 (100V)), 6605628 (TN; 5 CLR, FC500 (120V)) , 6605629 (TN; 5 CLR, FC500 (220V)), 6605630 (TN; 5 CLR, FC500 (240V)) with CXP Software Versions 2.0 and 2.1

Reason

Beckman Coulter has determined that under certain conditions a system error generated during a Pause and Rotate routine may result in a sample misidentification with a risk of reporting erroneous results.

Action

A Product Corrective Action letter was sent via US mail the week of May 22, 2006 to all FC500 customers. They were informed that under certain conditions a system error generated during a Pause and Rotate routine may result in a sample misidentification with a risk of reporting erroneous results. The letter includes the sequence in which the issue occurrs: 1) During the acquistion of a tube, the Pause and Rotate Button is selected. 2) A system error is generated during the Pause and Rotate routine (The error message table may be viewed by going to the CXP Help Menu, select the CXP Help Menu Item, then to Section 13.5). 3) An error message appears in the Cytometer Status, an audible alarm sounds, the run aborts, the carousel ejects, and the system automatically transitions from Verification to Run Initialization and then to Awaiting Sample 4) The Worklist is continued or a new Worklist is loaded and started. 5) After the Worklist is continued, or loaded and started, during the acquistion of a tube, the Pause and Rotate Button is selected. 6) The carousel Pauses but does not Rotate. 7) When the Start/Continue or Restart button is selected, the carousel may move to the wrong tube and aspirate sample resulting in a sample misidentification. If no system errors are generated during a Pause and Rotate routine, the system will function correctly. The letter also provides steps for immediate user preventive action if a system error is generated during the Pause and Rotate Routine to remedy the issue: 1) Exit CXP Software. 2) Power off the Cytometer using the FC OFF utility 3) Turn the Cytometer back on. 4) Recovery should be complete and the laboratory may resume normal usage of the system. Customers are asked to share the information with their laboratory staff and to retain this notification as part of their laboratory Quality System documentation and to complete and return the enclosed Fax Response Form for the firm''s records.

Distribution

Nationwide and Canada

Quantity

424