FDA Recall Open, Classified

Access Immunoassay Systems (1) Access 2 Immunoassay Analyzer (81600N); (2) Access 2 Immunoassay Analyzer, Refurbished (386220); (3) Access 2 Immunoassay Analyzer, Refurbished (A65531) ; and (4) Access 2 Section, DxC 600i Packaged(A25640).

Recall: Z-1257-2023 · Initiated January 26, 2023

Recall

Recall Number
Z-1257-2023
Event Number
91568
Firm
Beckman Coulter, Inc.
FEI Number
2122870
Product Code
JJE
Status
Open, Classified
Root Cause
Component change control
Initiated
January 26, 2023
Posted
March 15, 2023
Address
1000 Lake Hazeltine Dr, Chaska, MN, 55318-1037

Description

Access Immunoassay Systems (1) Access 2 Immunoassay Analyzer (81600N); (2) Access 2 Immunoassay Analyzer, Refurbished (386220); (3) Access 2 Immunoassay Analyzer, Refurbished (A65531) ; and (4) Access 2 Section, DxC 600i Packaged(A25640).

Reason

An increased incidence of damage to the nut band from a new vendor (Boamax) has led to an increased volume of complaints from consignees regarding wash arm motion failures, system check failures, and/or assay performance leading to a remote probability of delayed results or reporting erroneous patient results. The subject nut bands were found to not meet the firm's specifications for hardness and had the incorrect material grain direction. Affected instruments were manufactured from December 28, 2017 to October 19, 2022 or had their nut bands replaced during the same period.

Action

The recalling firm issued letters dated 1/19/2023 beginning 1/26/2023 via both email and first class mail. The letter explained the issue, which included photographs, impact, and the actions to be taken. The consignee was to inspect the wash arm lead screw nut band for damage following the instructions listed in Appendix B of the letter. This inspection was to be repeated during weekly maintenance until the affected nut band has been replaced. If the nut band appears to be intact and undamaged. the consignee may continue normal operation of the instrument. If Wash Arm Motion Errors are observed or if the nut band appears to be damaged in any way, specific instructions are provided. The recalling firm will replace the affected nut bands for all Access 2 Immunoassay Systems listed in Appendix A of the letter. A service representative will contact the consignee to schedule the wash arm lead screw nut band replacement. The information in the notification is to be shared with the laboratory staff and retained as part of the laboratory Quality System documentation. If the affected device has been forwarded to another laboratory, the consignee is requested to provide them a copy of the letter. Response to the notification is requested either electronically or manually using the enclosed response form within 10 days.

Distribution

Worldwide and Nationwide Distribution. There was government/military distribution. Foreign distribution was made to Algeria, Andorra, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Bonaire, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Ethiopia, France, Georgia, Germany, Ghana, Gibraltar, Greece, Guyana, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Kyrgyzstan, Lebanon, Libya, Lithuania, Macao, Malawi, Malaysia, Maldives, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, U.A.E, Uganda, Ukraine, United Kingdom, Uruguay, Vietnam, Yemen, Zambia, and Zimbabwe.

Quantity

4,137 devices (990 U.S. and 3,147 OUS)