FDA Recall
Terminated
CAROLINA(R) LIQUID CHEMISTRIES CORP. ETHA RGT. KIT, IVD, REF ST321
Recall: Z-1253-2019
·
Initiated January 16, 2019
Recall
- Recall Number
- Z-1253-2019
- Event Number
- 82017
- Firm
- Carolina Liquid Chemistries
- FEI Number
- 3009880320
- Product Code
- DIC
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- January 16, 2019
- Terminated
- February 2, 2021
- Address
- 313 Gallimore Dairy Rd, Greensboro, NC, 27409-9724
Description
CAROLINA(R) LIQUID CHEMISTRIES CORP. ETHA RGT. KIT, IVD, REF ST321
Reason
Supplier testing showed that the absorbance rate of the Ethyl Alcohol Assay did not meet internal quality control specifications. The defective lots will not function toward the end of shelf life.
Action
The firm initiated the recall by email on 01/16/2019. The notice requested that the consignee discard the product.
Distribution
NY. NH, ME, NC, TX, AR, SC, CT, IL, CA, NJ
Quantity
20 kits