FDA Recall Open, Classified

Atellica IM 1300 Analyzer - automated, immunoassay analyzer designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066001

Recall: Z-1252-2022 · Initiated April 13, 2022

Recall

Recall Number
Z-1252-2022
Event Number
90153
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2432235
Product Code
JLW
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
April 13, 2022
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

Atellica IM 1300 Analyzer - automated, immunoassay analyzer designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066001

Reason

Lots incompatible with Test Definition (TDef) Version 1.4, potentially lead to the generation of erroneous total T3 results

Action

Siemens issued an Urgent Medical Device Correction (UMDC) to US affected customers by FedEx beginning on 04/13/2022 and an Urgent Field Safety Notice (UFSN) was issued to all OUS. Letter states reason for recall, health risk and action to take: If your facility has multiple lots of T3 available for use and some kit lots are below lots ending in 244, follow the steps below to ensure that T3 results will be correct for all available lots until all kit lots available for use are kit lots ending in 244 or above. Ensure that all IM modules connected to an Atellica Solution have only 1 lot of T3 reagent (all IM modules have the same lot of T3 reagent packs) loaded onto every IM. For reagent kit lots below lots ending in 244: Rescan the Master Curve Card (MCC) for the T3 reagent lot that is on the Atellica IM before processing with that reagent lot, following instructions in the Atellica Online Help (SMN 11313586, section Scanning IM Master Curves and Assay Test Definitions 2D Barcodes ). Note: If the IM is rebooted while using kit lots below lots ending in 244, the MCC will need to be re-scanned before samples are processed. For reagent lots ending in 244 and above: Rescan the Master Curve Card (MCC) for the T3 reagent kit lot that is on the Atellica IM, following instructions in the Atellica Online Help (SMN 11313586, section Scanning IM Master Curves and Assay Test Definitions 2D Barcodes . Reboot all IM modules connected to Atellica Solution following instructions in the Atellica Online Help (SMN 11313586, section titled Shutting Down and Powering Off Analyzers within the System ) when changing from reagent kit lots below lots ending in 244 to a reagent pack associated with kit lots ending in 244 or above before processing with that reagent kit lot. Please review this letter with your Medical Director. Complete and return the Field Action Effectiveness Check Form attached to this letter within 30 days.

Distribution

Worldwide distribution - US Nationwide and the countries of Albania, Argentina, Austria, Bahamas, Bangladesh, Belgium, Brazil, Chile, China, Columbia, Croatia, Czech¿Republic, Denmark, Ecuador, Egypt, France, Germany, Greece, India, Israel, Italy, Japan, Korea, Latvia, Malaysia, Mexico, Netherlands, Paraguay, Peru, Philippines, Portugal, Romania, Russian¿Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., United Kingdom, Vietnam.

Quantity

1,566