FDA Recall
Terminated
SoftPath ASCII Software, used in pathology labs to track and report specimens and diagnostic results. Releases 1.2, 2.1, and 2.3
Recall: Z-1252-06
·
Initiated January 9, 2001
Recall
- Recall Number
- Z-1252-06
- Event Number
- 35375
- Firm
- SCC Soft Computer
- FEI Number
- 1000307651
- Product Code
- JQP
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 9, 2001
- Posted
- July 26, 2006
- Terminated
- July 19, 2010
- Address
- 34350 US Highway 19 N, Palm Harbor, FL, 34684-2149
Description
SoftPath ASCII Software, used in pathology labs to track and report specimens and diagnostic results. Releases 1.2, 2.1, and 2.3
Reason
The client had made a decision to use the SoftPath Module in a nonstandard manner. Under specific circumstances, incorrect text may appear on a patient report.
Action
The firm issued a software correction letter to its consignees 1/9/2001, advising them of the problem and supplied them with a software patch ICC number 3618 also referred to as release 3.0. The firm advised its clients that the installation of this software to their system was mandatory.
Distribution
Nationwide and Canada.
Quantity
171