FDA Recall Open, Classified

In Vitro Diagnostic (IVD) Panel BioFire BCID2 Panel - REF RFIT-ASY-0147

Recall: Z-1251-2023 · Initiated February 1, 2023

Recall

Recall Number
Z-1251-2023
Event Number
91647
Firm
BioFire Diagnostics, LLC
FEI Number
3002773840
Product Code
PEN
Status
Open, Classified
Root Cause
Environmental control
Initiated
February 1, 2023
Posted
March 15, 2023
Address
515 S Colorow Dr, Salt Lake City, UT, 84108-1248

Description

In Vitro Diagnostic (IVD) Panel BioFire BCID2 Panel - REF RFIT-ASY-0147

Reason

Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping delays.

Action

On 02/01/2023, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via email to customers informing them of a "Temperature Out of Range" issue with multiple products, as a result from Hurricane Ian (10/2022) Customers are Instructed to: Immediately examine their inventory for product identified in this recall. 1. Discontinue use and discard any remaining product in their possession that is subject to this recall. bioMerieux will replace the product at no charge in accordance with BIOFIRE s standard limited warranty. 2. Complete the accompanying Acknowledgment of Receipt Form and confirm the number of pouches scrapped (if any). Return the Acknowledgment of Receipt to bioMerieux so that bioMerieux may acknowledge their receipt of this notification. 3. If customers have further distributed this product, please identify any recipients, and notify them at once of this product recall. For questions or concerns, contact customer support department at [email protected] or via telephone by dialing +1.800.736.6354 and selecting option 5 for Product Technical Support

Distribution

U.S. Nationwide distribution in the states of FL and GA.

Quantity

2 kits