Elekta Leksell Stereotactic System, used for localization (spatial reference) for cranial surgery using X-ray or CT and MRI Image data.

Recall: Z-1251-2011 · Initiated October 2, 2009

Recall

Recall Number
Z-1251-2011
Event Number
57604
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
HAW
Status
Terminated
Root Cause
Labeling design
Initiated
October 2, 2009
Posted
February 15, 2011
Terminated
February 17, 2011
Address
4775 Peachtree Industrial Blvd, Bldg 300 # 300, Norcross, GA, 30092-3011

Description

Elekta Leksell Stereotactic System, used for localization (spatial reference) for cranial surgery using X-ray or CT and MRI Image data.

Reason

New protocols for MR sequences may result in higher RF energies deposited during MR scanning, generating heat in uninsulated fixation posts.

Action

Elekta sent Important Notice Affected Product: Fixation Posts, A331, dated June 22, 2009. The Important Notice will be added to the Customer User manual upon receipt.

Distribution

Nationwide

Quantity

379 units