FDA Recall
Terminated
Elekta Leksell Stereotactic System, used for localization (spatial reference) for cranial surgery using X-ray or CT and MRI Image data.
Recall: Z-1251-2011
·
Initiated October 2, 2009
Recall
- Recall Number
- Z-1251-2011
- Event Number
- 57604
- Firm
- Elekta, Inc.
- FEI Number
- 1037831
- Product Code
- HAW
- Status
- Terminated
- Root Cause
- Labeling design
- Initiated
- October 2, 2009
- Posted
- February 15, 2011
- Terminated
- February 17, 2011
- Address
- 4775 Peachtree Industrial Blvd, Bldg 300 # 300, Norcross, GA, 30092-3011
Description
Elekta Leksell Stereotactic System, used for localization (spatial reference) for cranial surgery using X-ray or CT and MRI Image data.
Reason
New protocols for MR sequences may result in higher RF energies deposited during MR scanning, generating heat in uninsulated fixation posts.
Action
Elekta sent Important Notice Affected Product: Fixation Posts, A331, dated June 22, 2009. The Important Notice will be added to the Customer User manual upon receipt.
Distribution
Nationwide
Quantity
379 units