Fetal Spiral Electrode, Model No. 9898 031 37631

Recall: Z-1250-2019 · Initiated January 25, 2019

Recall

Recall Number
Z-1250-2019
Event Number
82422
Firm
Philips North America, LLC
FEI Number
1218950
Product Code
HGP
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
January 25, 2019
Terminated
December 8, 2020
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

Fetal Spiral Electrode, Model No. 9898 031 37631

Reason

During use of the Philips FSE, it is possible for the metal electrode tip to break off and remain in the scalp of the newborn following labor and delivery.

Action

On January 25, 2019, the firm distributed Customer Information Medical Device Correction letters to affected customers. The letter informed customers that there was an increase in complaints regarding the Fetal Spiral Electrode tip breaking off, preventing easy removal from the scalp of the newborn infant. Customers were advised to do the following: * Visually examine the FSE upon removal from the newborn scalp to verify the tip is intact. If it is not intact, appropriate measures to locate and remove the missing part should be taken. * Upon receipt of this notice, clinicians should review the following information from the device's Instructions for Use: - Do not pull the Spiral Tip from the fetal skin. Do not pull the FSE wires apart. - Do not over-rotate Spiral Tip during attachment. - Inspect the Spiral Tip to ensure that it is still attached to the FSE Hub. If the tip has separated from the hub and remains embedded in the presenting part, remove it using aseptic technique. Customers were provided with a reply form and asked to complete the form and return it to Philips: 1-877-499-7223 (eFax) or email to [email protected]. If you need any further information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377, option 2.

Distribution

Distributed nationwide within US. Worldwide international distribution.

Quantity

1,971,925 (78,877 boxes)