Medtronic EnTrust, Dual Chamber Implantable, Cardioverter Defibrillator with Ventricular Therapies, models D153DRG and D154DRG. The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment for automated treatment of life-threatening ventricular arrhythmias. Sterile. Medtronic, Inc., Minneapolis, MN 55432 USA.
Recall
- Recall Number
- Z-1250-2012
- Event Number
- 61356
- Firm
- Medtronic Inc. Cardiac Rhythm Disease Management
- FEI Number
- 2182208
- Product Code
- LWS
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- March 6, 2012
- Posted
- March 22, 2012
- Terminated
- July 27, 2012
- Address
- 8200 Coral Sea St. N.E., Saint Paul, MN, 55112
Description
Medtronic EnTrust, Dual Chamber Implantable, Cardioverter Defibrillator with Ventricular Therapies, models D153DRG and D154DRG. The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment for automated treatment of life-threatening ventricular arrhythmias. Sterile. Medtronic, Inc., Minneapolis, MN 55432 USA.
A small percentage of EnTrust ICDs may not meet expected longevity or provide at least three months of device operation between the Elective Replacement Indicator (ERI) and End of Life (EOL) due to a more-rapid-than-expected drop in battery voltage.
An "Urgent Medical Device Correction" letter was sent to physicians beginning 3/6/2012. The letter described the issue and provided patient management recommendations. Additionally the letter stated that ongoing updates to the product performance report will be posted on a Medtronic website.
Worldwide Distribution -- USA, and the countries of: Argentina, Aruba, Australia, Austria, Bahamas, Barbados, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jamaica, Jordan, Kuwait, Lebanon, Macedonia, Martinique, Morocco, Netherlands, Netherlands Antilles, New Zealand, Norway, Oman, Peru, Poland, Portugal, Qatar, Reunion, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
68,804 devices worldwide