FDA Recall Terminated

Hamilton brand RAPHAEL Silver Ventilator (Software Version 2.2xS) Product is distributed by Hamilton Medical, Inc., 4990 Energy Way, Reno, NV 89502

Recall: Z-1250-06 · Initiated May 24, 2006

Recall

Recall Number
Z-1250-06
Event Number
35592
Firm
Hamilton Medical Inc
FEI Number
2937708
Product Code
CBK
Status
Terminated
Root Cause
Other
Initiated
May 24, 2006
Posted
July 22, 2006
Terminated
December 21, 2006
Address
4990 Energy Way P O Box 30008, Reno, NV, 89520-3008

Description

Hamilton brand RAPHAEL Silver Ventilator (Software Version 2.2xS) Product is distributed by Hamilton Medical, Inc., 4990 Energy Way, Reno, NV 89502

Reason

ALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.

Action

On May 24, 2006 the manufacturer initiated this recall. The Recalling establishment/US Initial Importer/Distributor-received the manufacturer''s notice on May 31, 2006 and initiated their recall on June 1, 2006. The US distributor has issued a Medical Device Field Correction.

Distribution

Worldwide

Quantity

2,385 for all models