FDA Recall Open, Classified

Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney, is the principal factor regulating red blood cell production (erythropoiesis) in mammals. Renal production of EPO is regulated by changes in oxygen availability. Under conditions of hypoxia, the level of EPO in the circulation increases and this leads to increased production of red blood cells.

Recall: Z-1249-2025 · Initiated February 3, 2025

Recall

Recall Number
Z-1249-2025
Event Number
96255
Firm
Beckman Coulter, Inc.
FEI Number
2122870
Product Code
GGT
Status
Open, Classified
Root Cause
Unknown/Undetermined by firm
Initiated
February 3, 2025
Posted
February 28, 2025
Address
1000 Lake Hazeltine Dr, Chaska, MN, 55318-1037

Description

Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney, is the principal factor regulating red blood cell production (erythropoiesis) in mammals. Renal production of EPO is regulated by changes in oxygen availability. Under conditions of hypoxia, the level of EPO in the circulation increases and this leads to increased production of red blood cells.

Reason

Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased results, which may lead to improper diagnosis or repeat testing.

Action

Firm began notifying customers February 6, 2025. Customers were instructed to discontinue using and discard all remaining reagent packs of the Access EPO reagent lot identified. Customers are advised to share this recall with their medical director and perform a retrospective review of patient test results generated with this lot and assess the need for retesting.

Distribution

Domestic distribution to AL AR AZ CA CT FL ID IL IN KS MA MI NC NE NJ NY OH OR PA TN TX UT WA WV. International distribution to Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Korea, Republic of, Netherlands, Philippines, Russian Federation, Saudi Arabia, Slovakia, Spain, United Kingdom of Great Britain and Northern Ireland

Quantity

498 US; 1576 OUS