Getinge CM320 WUWD Series Washer Disinfector-For use as a multi-chamber washer-disinfector for the washing, disinfecting and drying surgical instruments, Model: CM320 WUWD

Recall: Z-1249-2022 · Initiated April 13, 2022

Recall

Recall Number
Z-1249-2022
Event Number
90146
Firm
Getinge Usa Sales Inc
FEI Number
3013876692
Product Code
MEC
Status
Open, Classified
Root Cause
Process control
Initiated
April 13, 2022
Address
1 Geoffrey Way, Wayne, NJ, 07470-2035

Description

Getinge CM320 WUWD Series Washer Disinfector-For use as a multi-chamber washer-disinfector for the washing, disinfecting and drying surgical instruments, Model: CM320 WUWD

Reason

Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall injury. May lead to a patient procedural delay and/or infection

Action

Getinge issued to U.S. consignees Medical Device Correction Letter dated April 13, 2022 via FedEx delivery. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any Getinge CM320-series Washer Disinfectors with the REF number/serial numbers listed on page 1. You can continue to use Getinge CM320-series Washer Disinfectors until the corrective action is performed; there are no precautionary actions necessary. 2. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 5) to acknowledge that you have received this notification. A Getinge service technician will review the Installation Checklist and verify that each of the steps were performed to ensure continual safe operation of your cart washer(s). This review and verification will be performed. Questions contact your Getinge representative or Technical Support please call (888) 943-8872 (options 4,2,1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). Technical Support can also be reached via email at [email protected].

Distribution

US Nationwide distribution in the states of MD, MN, ND, NE, NY, PA, VA.

Quantity

11 units