FDA Recall Terminated

cobas e 801 immunoassay analyzer

Recall: Z-1249-2019 · Initiated March 15, 2019

Recall

Recall Number
Z-1249-2019
Event Number
82394
Firm
Roche Diagnostics Corporation
FEI Number
1823260
Product Code
JJE
Status
Terminated
Root Cause
Device Design
Initiated
March 15, 2019
Terminated
April 23, 2020
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

cobas e 801 immunoassay analyzer

Reason

The firm has confirmed complaints regarding signal drops on the cobas e 801 module after a ProCell II M bottle changeover, which may lead to incorrect medical decisions with respect to diagnostics and patient treatment.

Action

On March 15, 2019, the firm notified affected customers of the recall via an Urgent Medical Device Correction letter. Customers were informed of the potential problem of signal drops when the standby bottle flowpath is covered with a biofilm. Customers were asked to do the following: " Follow the Workaround section to determine if your cobas e 801 module is potentially affected and perform the steps if applicable. " If your cobas e 801 module is potentially affected based on the information in the Workaround section of this UMDC, contact the Roche Support Network Customer Support Center at 1-800-428-2336 to schedule the ProCell II M flowpath decontamination procedure. The ProCell II M flowpath decontamination procedure should be performed every 4 weeks until your cobas e 801 module is switched to the improved ProCell II M formulation. " Identify and retest potentially affected patient samples using the Retrospectively Identify and Correct Potentially Affected Patient Sample Results section of this UMDC. " Complete the attached fax form and fax it to 1-833-254-2597 or email it to [email protected].

Distribution

Distributed to accounts in AL AZ CA CT FL HI IA IL IN KY LA MA MD ME MI MN MO NC NH NJ NY OH OR PA PR SC TN TX VA WA WI

Quantity

149