FDA Recall Terminated

Medtronic EnTrust, Single Chamber Implantable, Cardioverter Defibrillator with Ventricular Therapies, models D153VRC and D154VRC. The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment for automated treatment of life-threatening ventricular arrhythmias. Sterile. Medtronic, Inc., Minneapolis, MN 55432 USA.

Recall: Z-1249-2012 · Initiated March 6, 2012

Recall

Recall Number
Z-1249-2012
Event Number
61356
Firm
Medtronic Inc. Cardiac Rhythm Disease Management
FEI Number
2182208
Product Code
LWS
Status
Terminated
Root Cause
Device Design
Initiated
March 6, 2012
Posted
March 22, 2012
Terminated
July 27, 2012
Address
8200 Coral Sea St. N.E., Saint Paul, MN, 55112

Description

Medtronic EnTrust, Single Chamber Implantable, Cardioverter Defibrillator with Ventricular Therapies, models D153VRC and D154VRC. The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment for automated treatment of life-threatening ventricular arrhythmias. Sterile. Medtronic, Inc., Minneapolis, MN 55432 USA.

Reason

A small percentage of EnTrust ICDs may not meet expected longevity or provide at least three months of device operation between the Elective Replacement Indicator (ERI) and End of Life (EOL) due to a more-rapid-than-expected drop in battery voltage.

Action

An "Urgent Medical Device Correction" letter was sent to physicians beginning 3/6/2012. The letter described the issue and provided patient management recommendations. Additionally the letter stated that ongoing updates to the product performance report will be posted on a Medtronic website.

Distribution

Worldwide Distribution -- USA, and the countries of: Argentina, Aruba, Australia, Austria, Bahamas, Barbados, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jamaica, Jordan, Kuwait, Lebanon, Macedonia, Martinique, Morocco, Netherlands, Netherlands Antilles, New Zealand, Norway, Oman, Peru, Poland, Portugal, Qatar, Reunion, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Quantity

68,804 devices worldwide