FDA Recall Open, Classified

ASP BIOTRACE Auto Read 60 Steam BI Process Challenge Device, REF 73135.

Recall: Z-1248-2025 · Initiated January 15, 2025

Recall

Recall Number
Z-1248-2025
Event Number
96176
FEI Number
3013145340
Product Code
FRC
Status
Open, Classified
Root Cause
Labeling False and Misleading
Initiated
January 15, 2025
Posted
February 27, 2025
Address
TERRAGENE S.A. Ruta Nacional N 9, Km 280, Cp 2130 Parque Industrial Micropi Alvear Argentina

Description

ASP BIOTRACE Auto Read 60 Steam BI Process Challenge Device, REF 73135.

Reason

Error in the IFU lists an incorrect incubation time on one occasion of 20 minutes rather than the correct incubation time of 60 minutes.

Action

The recalling firm issued an email on 1/15/2025 to the main office of the distributor with an attached Distributor Customer Communication letter dated 1/15/2025, an Acknowledgement and Receipt Form, an updated IFU (Rev.B R01), and a User Letter for the distributor to issue to their customers. The letter to the distributor explains that all product is affected that contains version Rev.A R0 of the IFU because a discrepancy was identified in that version. The Distributor is to review their inventory records for the affected product, identify customers that received or may have received the devices and report it to the recalling firm through the provided Acknowledgement and Receipt Form via email. The Distributor is informed the product can be used according to the enclosed updated IFU (Rev.B R01). The Distributor was requested to notify their customer by using the enclosed User Letter and include the updated IFU with the User Letter. The Distributor is also informed they can continue distribution of their inventory by sending the digital user letter and the updated digital IFU with each delivery. An updated printed IFU will also be made available to users upon request. The User Letter was similar to the Distributor letter except that it instructs the user to discard older version of the IFU (Rev.A R0 and use the product according to the IFU attached to the letter and available on the firm's website. an Acknowledgement and Receipt Form was also enclosed for completion and return.

Distribution

US Nationwide distribution in the states of CA and KY.

Quantity

914 devices in the U.S.